Multiparameter Photometer & pH Meter for Water Conditioning- HI83308-02
HI83308 is a compact, multiparameter photometer for use in water conditioning applications.
The meter is one of the most advanced photometers available with a innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette.
This meter has 23 different programmed methods measuring 15 key water quality parameters and also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves. Water conditioning parameters include fluoride, chlorine, ammonia, manganese, copper, iron, silver and zinc.
To save valuable laboratory benchtop space, the HI83308 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.
pH Range | Photometer: 6.5 to 8.5 pH pH electrode: -2.00 to 16.00 pH |
---|---|
pH Resolution | Photometer: 0.1 pH pH electrode: 0.01 pH |
pH Accuracy | Photometer: 0.1 pH pH electrode: 0.01 pH |
pH Calibration | Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
pH Temperature Compensation | Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used |
pH CAL Check (electrode diagnostics) | clean electrode and check buffer/check probe displayed during calibration |
pH Method | Photometer: phenol red |
pH-mV Range | 1000 mV |
pH-mV Resolution | 0.1 mV |
pH-mV Accuracy | 0.2 mV |
Oxygen, Dissolved Range | 0.0 to 10.0 mg/L (as O2) |
---|---|
Oxygen, Dissolved Resolution | 0.1 mg/L |
Oxygen, Dissolved Accuracy | 0.4 mg/L 3% of reading |
Oxygen, Dissolved Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method |
Absorbance Range | 0.000 to 4.000 Abs |
---|---|
Absorbance Resolution | 0.001 Abs |
Absorbance Accuracy | +/-0.003Abs @ 1.000 Abs |
Ammonia Range | Low Range: 0.00 to 3.00 mg/L Medium Range: 0.00 to 10.00 mg/L High Range: 0.0 to 100.0 mg/L (all as NH3-N) |
---|---|
Ammonia Resolution | 0.01 mg/L; 0.1 mg/L |
Ammonia Accuracy | Low Range: 0.04 mg/L 4% of reading Medium Range: 0.05 mg/L 5% of reading High range: 0.5 mg/L 5% of reading |
Ammonia Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method |
Free Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) |
---|---|
Free Chlorine Resolution | 0.01 mg/L |
Free Chlorine Accuracy | 0.03 mg/L 3% of reading |
Total Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) |
Total Chlorine Resolution | 0.01 mg/L |
Total Chlorine Accuracy | 0.03 mg/L 3% of reading |
Chlorine Method | Adaptation of the EPA 330.5 DPD method |
Copper Range | Low Range: 0.000 to 1.500 mg/L (as Cu2+) High range: 0.00 to 5.00 mg/L (as Cu2+) |
---|---|
Copper Resolution | 0.001 mg/L; 0.01 mg/L |
Copper Accuracy | Low Range: 0.01 mg/L 5% of reading High Range 0.02 mg/L 4% of reading |
Copper Method | Adaptation of the EPA bicinchoninate method |
Fluoride Range | 0.00 to 2.00 mg/L (as F-) |
---|---|
Fluoride Resolution | 0.01 mg/L |