Main centres: | 1-3 business days |
Regional areas: | 3-4 business days |
Remote areas: | 3-5 business days |
Designed for an out-of hospital and hospital use.
Lightweight, ergonomic, attractive and robust portable device. Rescue Life 9 provides complete monitoring of parameters like 12 channels ECG, HR, SpO2, NIBP, Temperature and EtCO2. Synchronized cardioversion and external pacemaker in fixed or demand mode. Three operating modes: manual, AED, Advisory. Capnography waveform. Auto-test mode verifies device functions, battery status and charge-discharge system with real time data management and report printing. The IPX4 permits operations under adverse climatic situations.
Temperature (standard) and EtCO2 (optional extra not included in the unit)
It includes: temperature and EtCO2 configuration.
Optional Extra EtCO2 module can be added anytime.
Dual battery: one included, one optional
The powerful battery is up to 5 hours continuous monitoring. Second battery improves working time operation to 10 hours and avoid process disruption. Batteries can be easily replaced by the operator.
High resolution TFT LCD colour 8.4'' display Intelligent Operator Support
Display shows visual messages, different colours for each function, clear voice support during difficult situations and simple menu with dedicated functions keys.
ECG Analysis in two modes
ECG signal can be filtered through diagnostic mode, which collects ECG traces for qualified personnel interpretation and diagnosis; monitoring mode, which removes noises caused by patient's movements or body physiological process.
AED & Advisory mode
AED mode: operators are supported by voices and messages in CPR and shock release. Advisory mode automatically and continuously analyzes ECG rhythm and the connection of cable and pads on patient, and tells if the shock is necessary.
Memory and Data Management + Software
Recorded ECG data and events can be transferred and managed on PC through a software which can be freely downloaded.
Other features:
- alarms for each single function
- automatic report before and after the shock release
- ergonomic reusable pads for manual defibrillation, disposable pads for AED defibrillation.
Technical Specifications
General requirements
Dimension: 335x225x360 mm
Weight: 6.5 kg with dual batteries
Electrical: AC 100-240 VAC, 50/60 Hz
Rechargeable Lithium-ion batteries: 15.0 V - 3.2 Ah
Battery Operation Time:
>5 hours, with 1 battery
>10 hours with 2 batteries
Charge Time: Max. 3 hours
Capacity: 150 shocks with 1 battery or 300 shocks with 2 batteries at 230J fully charged
Software: Voice and texts multi language
Memory: Internal flash for ECG data, events and usage review
Defibrillation
Waveform: Biphasic Truncated Exponential (BTE), HiCAP technology (Large storage capacitor)
Impedance: C ompensated, range 25-200 Ohm
Charge time: 6 sec for 230 J with battery fully charged
AED mode
Energy: Fixed energy at 200 J
Protocol: ERC 2015 CPR Guidelines with voice and text prompts
Shockable rhythms: VF with amplitude > 0.15 mV and VT with rhythm >150 bpm
Sensitivity: S hockable rhythm -VF > 95%, -VT > 75%
Specificity: According to ERC 2015,
Normal sinusoidal rhythm > 99%
Asystole and other non-shockable rhythms > 99%
Manual mode
Energy range: 1-230 J (from 1-10 J in 1 J steps; from 10 to 230 J in 10 J steps)
Commands: Multifunction trim knob
Charge and shock button in the front panel
Operating mode: EC G R wave synchro or async mode
Indicators: B attery and main led indicators
Display
Type: High contrast TFT colour LCD 8" (800x600 dots)
Information: ECG traces, SpO2, NIBP, EtCO2, TEMP, Heart Rate indicator, filters, alarms and batteries status
Display mode: Can select 3, 6, 12 ECG leads
Ecg monitoring
Bandwidth:
0.05 - 120 Hz (-3 db) in diagnostic mode
0.6 - 40 Hz (-3 db) in monitoring mode
CMRR: > 98 dbv Input impedance: > 50 Mohm
Input:
Patient cable 5 leads (I, II, III, aVL, aVF, aVR, V) - standard
Patient cable 10 leads (I, II, III, aVL, aVF, aVR, V1..V6) - optional
Speed: 5, 10, 25, 50 mm/sec
Filters: 50/60 Hz, EMG Filter, base line
HR range: 20-300 bpm (~2%), digital readout on the display
Alarm: HR max settable 250 bpm; HR min 20 bpm
SpO2 (only 55505-55512)
SpO2 range: 0-100% HR range: 30-250 bpm
Accuracy: 70-100% ~2% for adults with finger clip sensor
Alarm: Adjustable min 50%
Oscillometric NIBP (only 55510-55512)
Accuracy: Meets ANSI/AAMI SP10-2002, EN 1060-4
Application: Adult / Paediatric / Neonatal (mmHg)
Systolic Range: 40-260 40-160 40-130
Range MAP: 26-220 26-133 26-110
Diastolic Range: 20-200 20-120 20-100
PR range: 30-220 bpm 2% or 3 bpm
Operating Modes: Manual, Long Term Automatic, Stat, Service
Auto Interval Periods: 1,2,3,4,5,10,15,30,60 and 90 minutes
Integrated thermal 3 channels printer
Paper speed 5, 10, 25, 50 mm/sec Paper width: 58 mm
Mode: Manual, automatic (12" pre and post shock recording)
External pacemaker
Type: Rectangular wave
Operating mode: Fixed on demand
Pulse rate: 30 to 250 ppm, adj. in steps of 5 ppm
Impulse duration: 22.5 ms
Pulse current: 0 to 180 mA adj. in steps of 5mA
Amplitude: Max 200 V
Mainstream technology Entidal CO2 (optional extra not included with the unit)
Suitable for both intubated and non-intubated patients
Measurement range: 0 to 150mmHg 2%
Respiratory rate: 0 to 150 bpm
Temperature
Probe type: YSI400 compatible
Range: 0 to 50 C
Resolution: 0.1 C
Standard
IEC 60601-2-4; IEC 60601-1; IEC 60601-1-2; IEC 63204, IEC 62366-1
CE Mark EEC 93/42: Medical device, Class IIb
Isolation: ECG type CF, Defibrillation type BF
Water proof: Class IP54
Brochure